Hemavet 950

New, Fast, Veterinary Multi-species Hematology System


Total Quality Results Right in your Office
 
Features :
18 parameters plus platelets
Fast : 26 samples per hour
Complete 5 part WBC differential
Designed specifically for veterinary medicine
Patented Expectation Maximisation software
Fully automated, whole blood open vials
Simple to operate
Low maintenance, due to Focused Flow® technology
Optimized Reagents and Controls
Efficient, cost effective patient management

Parameters :
 

WBC, NE#, NE%, LY#, MO#, MO%, EO%, BA#, BA%
RBC, Hb, HCT, MCV, MCH, MCHC, RDW
plus additional flagging for NRBC#, NRBC%, Reticulocytes#, Reticulocytes%note2
Platelet Count, MPV
3D cytograms : WBC subpopulations, RBCs, PLTs

Species :
Up to Six Dedicated Pre-programmed Species Keys.
(Additional species can be run on 'OTHER' key)note1
Available Species :
  Bison, Calfnote2, Camel², Cat, Cow, Deer², Dog, Elephant², Elk4, Ferret², Foal², Goat², Guinea Pig², Horse, Llama², Monkey, Mouse, Pig4, Pre-Dilute Mouse, Pre-Dilute Rat, Rabbit4, Rat, Sheep4
Sample Delivery :
Automated open vial sample delivery. Automated probe wiping. One touch walk-away sample processing.
Sample Volume :
20 µL whole blood  
Sample Throughput :
  Approximately 26 samples per hour  
Data Input / Output :

Touch Keypad with Six Dedicated Species Keys and one 'OTHER' key
LCD display
Customised report (8.5" x 11") with exclusive 3-D cytograms of WBC and RBC subpopulations
RS232 port for external computer interface

Patient Storage :
Internal data Storage for last 50 samples run.
Flagging :
Automatic distribution and morphological flagging.
Hematologic abnormality charts indicating Mild, Moderate and Severe conditions.
Reagent Kit :
  Reagent Kit optimized for the Hemavet 950. This kit yields approximately 135 billable tests per reagent kit when running 15 samples per day.
Controls :
  Dog, Cat, Mouse and Horse Controls are available.
Calibration :
  Drew Scientific recommends running a control each day. The calibration should be reviewed every six months. Adjust as necessary. Review after replacing worn or damaged components

Performance :
Precision :
Level Units CV
WBC 10.0 x 10³ cells/µL < 3.0%
RBC 7.0 x 106 cells/µL < 3.0%
Hb 14 g/dL < 2.0 %
MCV 60 fL < 1.0 %
PLT 500 x 10³ cells/µL < 5.0 %
NE# 6.0 x 10³ cells/µL < 6.0 %
LY# 6.0 x 10³ cells/µL < 6.0 %
EO# 2.0 x 10³ cells/µL < 12.0 %
Linearity :
  Range Units  
WBC 0.1 to 200.0 x 10³ cells/µL  
RBC 0.01 to 20.0 x 106 cells/µL  
Hb 0.1 to 35 g/dL  
MCV 30 to 300 fLnote3  
PLT 1 to 4000 x 10³ cells/µL  
NE% 0.1 to 99.9%    
LY% 0.1 to 99.9%    
EO% 0.1 to 99.9%    

 
Note 1 Differential parameters and flagging not available with "OTHER" key
Note 2 Differential parameters and abnormal flagging not available with all Species
Note 3 Extended MCV linearity range (10 to 600 fL) available for Exotic Species Packages, eg Goat, Llama
Note 4 RBC and PLT Cytograms not available with all Species

 
Temperature
Operation: 62ºF to 90ºF (17ºC to 32ºC)
Voltage: 95 to 132 vAC or 180 to 264 vAC
47 to 63 Hz
Dimensions : Height : 12"     (30.5 cm)
Width :  10.5"  (27cm)
Depth :  20"     (51cm)
Weight : 25.5 lbs  (11.6 Kg)