5 Part Differential Automatic Hematology Analyzer |
|
|
| Features : | ||
|
|
22 parameters, 2 scattergrams and 2 histograms | |
|
|
Proprietary laser technology, 5 part differential | |
|
|
Impedance technology for cell counting | |
|
|
State of the art patented platelet counting technique | |
|
|
70 samples per hour | |
|
|
Sample size flexibility satisfies a wide range of end user requirements | |
|
|
SVGA color monitor with scattergrams and histograms on screen | |
|
|
User friendly operator interface (keyboard and mouse) | |
|
|
Data management stores up to 100,000 sample records including scattergrams and histograms | |
|
|
Built in QC package (Levey-Jennings, X-B) | |
|
|
Bidirectional port for communications with LIS | |
|
|
Modern electronics with industrial grade processors for durability and serviceability | |
|
Optional
Features :
|
||
|
|
Automatic 30 tube sampler/mixer with optional bar code sample identification (as shown) | |
|
|
||
| Parameters : | ||
| WBC,
Lymp#, Gran#, Mono#, Eos#, Baso#, Lymp%, Gran%, Mono%, Eos%, Baso%, RBC, Hgb, Hct, MCV, MCH, MCHC, RDW, Plt, MPV, 2 x WBC scattergrams Plt histogram, RBC histogram |
||
| Principle : | ||
| 4
x dimensions optical scattering for WBC differential Electrical resistance for counting (WBC, RBC, Plt) and sizing (Plt, RBC) Optical absorbance of cyanmethemaglobin for hemaglobin |
||
| Sample volumes : | ||
| Direct
mode Sample saver mode Automatic sampler |
180
µL whole blood 80 µL whole blood 180 µL whole blood |
|
| Dilution System : | ||
| High precision ceramic shear valve and computer controlled diluent dispenser. Automatic rinse inside and outside of the sample probe prevents carryover and reduces risk of transmission of blood borne pathogen to operators. | ||
| Sensors : | ||
| Flow cytometer with solid state sensors or photomultiplier for WBC differential. | ||
| WBC aperture | 100 µm dia x 75 µm length | |
| RBC/Plt aperture | 78 µm dia x 55 µm length | |
| The Plt/RBC channel features the patented "von Behrens" silencer, which eliminates RBC recirculation that interferes with low Plt counts in conventional aperture designs. | ||
| Throughput : | ||
| Minimum 70 samples per hour in direct mode. | ||
|
|
| Performance : |
| Precision : | |
| Level | Units | CV | |
| WBC | 8.0 | x 10³ cells/µL | < 1.5 % |
| RBC | 8.0 | x 106 cells/µL | < 1.0 % |
| Hgb | 16 | g/dl | < 1.0 % |
| MCV | 90 | fl | < 1.0 % |
| Plt | 250 | x 10³ cells/µL | < 3.0 % |
| Linearity : | |||
| Range | Units | Linearity
(whichever greater) |
|
| WBC | 0.1 to 99.9 | x 10³ cells/µL | ± 0.2 or ± 2% |
| RBC | 0.02 to 9.99 | x 106 cells/µL | ± 0.05 or ± 2% |
| Hgb | 0.0 to 30 | g/dl | ± 0.2 or ± 2% |
| MCV | 40 to 150 | fl | ± 0.2 or ± 2% |
| Plt | 10 to 1000 | x 10³ cells/µL | ± 10 or ± 5.0% |
| Plt | 10 to 2000 | x 10³ cells/µL | ± 15 or ± 7% |
|
|
|
| Operating system : | |
| Microsoft Windows 2000 | |
| Software Package : | |
|
Patient identification
: PID 12 digits, name, first name, sex, DoB, comments |
|
| Display : | |
| SVGA monitor | |
| User Interface : | |
| 101 key alpha numeric keyboard, mouse and light pen (optional bar code reader) | |
| Ports : | |
| Two RS-232 serial ports and one parallel printer port (Centronics) | |
| Printer (optional) : | |
| External : Windows 2000 supported, colour or B/W | |
|
|
|
|
Temperature
|
Humidity | |
| Operation: | 15ºC to 32ºC | 10% - 90% non-condensing |
| Voltage: | 90
to 250 VAC 47 to 63 Hz (no manual adjustment) |
|
| Dimensions (analyzer) | 42cm (h) x 58cm (w) x 50 cm (d) | |
| Weight (analyzer) |
32.5 Kg |
|
|