Excell 22 Excell 22

5 Part Differential Automatic Hematology Analyzer


Features :
22 parameters, 2 scattergrams and 2 histograms
Proprietary laser technology, 5 part differential
Impedance technology for cell counting
State of the art patented platelet counting technique
70 samples per hour
Sample size flexibility satisfies a wide range of end user requirements
SVGA color monitor with scattergrams and histograms on screen
User friendly operator interface (keyboard and mouse) 
Data management stores up to 100,000 sample records including scattergrams and histograms
Built in QC package (Levey-Jennings, X-B)
Bidirectional port for communications with LIS
Modern electronics with industrial grade processors for durability and serviceability
Optional Features :
Automatic 30 tube sampler/mixer with optional bar code sample identification (as shown)

Parameters :
  WBC, Lymp#, Gran#, Mono#, Eos#, Baso#,
           Lymp%, Gran%, Mono%, Eos%, Baso%,
RBC, Hgb, Hct, MCV, MCH, MCHC, RDW,
Plt, MPV,
2 x WBC scattergrams
Plt histogram, RBC histogram
Principle :
  4 x dimensions optical scattering for WBC differential
Electrical resistance for counting (WBC, RBC, Plt) and sizing (Plt, RBC)
Optical absorbance of cyanmethemaglobin for hemaglobin
Sample volumes :
  Direct mode
Sample saver mode
Automatic sampler
180 µL whole blood
80 µL whole blood
180 µL whole blood
Dilution System :
  High precision ceramic shear valve and computer controlled diluent dispenser. Automatic rinse inside and outside of the sample probe prevents carryover and reduces risk of transmission of blood borne pathogen to operators.
Sensors :
  Flow cytometer with solid state sensors or photomultiplier for WBC differential.
  WBC aperture 100 µm dia x 75 µm length
  RBC/Plt aperture 78 µm dia x 55 µm length
  The Plt/RBC channel features the patented "von Behrens" silencer, which eliminates RBC recirculation that interferes with low Plt counts in conventional aperture designs.
Throughput :
  Minimum 70 samples per hour in direct mode.

Performance :
   
Precision :
  Level Units CV
WBC 8.0 x 10³ cells/µL < 1.5 %
RBC 8.0 x 106 cells/µL < 1.0 %
Hgb 16 g/dl < 1.0 %
MCV 90 fl < 1.0 % 
Plt 250 x 10³ cells/µL < 3.0 %
Linearity :
  Range Units Linearity
(whichever greater)
WBC 0.1 to 99.9 x 10³ cells/µL ± 0.2 or ± 2%
RBC 0.02 to 9.99 x 106 cells/µL ± 0.05 or ± 2%
Hgb 0.0 to 30 g/dl ± 0.2 or ± 2%
MCV 40 to 150 fl ± 0.2 or ± 2%
Plt 10 to 1000 x 10³ cells/µL ± 10 or ± 5.0%
Plt 10 to 2000 x 10³ cells/µL ± 15 or ± 7%

Operating system :
  Microsoft Windows 2000™
Software Package :
 

Patient identification : PID 12 digits, name, first name, sex, DoB, comments
Patient data storage : 100,000 results with scattergrams & histograms
User selectable report format
ASTM or HL7 report format
Automatic or manual calibration
Selectable units : MKS, SI, SI modified
Patient normal range : 8 sets, user modified
Patient action limits : user defined
Multi-language capability
QC assay files : 12 lots, 3 levels each.
Assays are loaded via keyboard, modem or disk
Printed log of QC data with automatic calculation of mean, SD, CV
Monthly Levey-Jennings plot of QC data, X-B, Bull's moving average, Westgard rules
QC results transfer for peer to peer analysis via modem or disk
Diagnostic alerts to assist operator
Maintenance log
Errors log

Display :
  SVGA monitor
User Interface :
  101 key alpha numeric keyboard, mouse and light pen (optional bar code reader)
Ports :
  Two RS-232 serial ports and one parallel printer port (Centronics)
Printer (optional) :
  External : Windows 2000 supported, colour or B/W

   
 
Temperature
Humidity
Operation: 15ºC to 32ºC 10% - 90% non-condensing
Voltage: 90 to 250 VAC
47 to 63 Hz
(no manual adjustment)
 
Dimensions (analyzer) 42cm (h) x 58cm (w) x 50 cm (d)  
Weight
(analyzer)
32.5 Kg