HbA1c measurement |
|
|
|
|
HbA1c result in 5 minutes |
|
|
Immediate labile fraction removal using the DS5 dispenser |
|
|
Simple to use with icon driven touch screen |
|
|
15 position autosampler |
|
|
Detects presence of HbS or HbC |
|
|
|
|
|
Intended Use The DS5 instrument, dispenser and associated reagent kit are intended for the measurement of HbA1c in blood samples. The system's small size and ease of use makes it ideal for main laboratory, clinic or satellite laboratory settings Scientific principle The DS5 uses cation exchange chromatography in conjunction with gradient elution to separate human haemoglobin subtypes and variants from haemolysed whole blood. The separated haemoglobin fractions are monitored by means of absorption of light and the chromatogram obtained is recorded and stored by the internal computer. A software program in the instrument performs the analysis of the chromatogram and generates a results report on the integral thermal printer. Performance Data |
| Linearity | |
| The DS5 produces linear results between 4.15 and 14.9% HbA1c. | |
| Within batch precision : |
| HbA1c | HbA1c | ||
| Number | 10 | 10 | |
| Mean value | 5.0 | 13.2 | |
| SD | 0.11 | 0.26 | |
| CV | 2.3% | 2.0% |
| Between batch precision : |
| HbA1c | HbA1c | ||
| Number | 10 | 10 | |
| Mean value | 4.3 | 7.0 | |
| SD | 0.08 | 0.27 | |
| CV | 1.9% | 3.9% |
|
|
| Supporting Products |
|
Reagent Kit (300 Tests) |
| Reagent Kit (75 Tests) |
| Bi-level HbA1c QC Kit (3 x 0.3 ml level1, 3 x 0.3 ml level2) |
| Technical Data |
| Data Storage |
| Storage for 15 chromatograms, results and summary, RS232 Unidirectional Serial Interface Port. |
|
Temperature
|
Humidity | |
| Operation: | 15ºC to 35ºC | 10% - 80% non-condensing |
| Transportation: | -5ºC to 40ºC | 5% - 80% non-condensing |
| Voltage: | 110/115V or 220/240V:50/60Hz | |
| Dimension | 27cm (w) x 41 cm (d)x 35 cm (h) | |
| Weight | 20 Kg |
|
|