DS5

HbA1c measurement


HbA1c result in 5 minutes
Immediate labile fraction removal using the DS5 dispenser
Simple to use with icon driven touch screen
15 position autosampler
Detects presence of HbS or HbC
 

Intended Use

The DS5 instrument, dispenser and associated reagent kit are intended for the measurement of HbA1c in blood samples. The system's small size and ease of use makes it ideal for main laboratory, clinic or satellite laboratory settings

Scientific principle

The DS5 uses cation exchange chromatography in conjunction with gradient elution to separate human haemoglobin subtypes and variants from haemolysed whole blood. The separated haemoglobin fractions are monitored by means of absorption of light and the chromatogram obtained is recorded and stored by the internal computer. A software program in the instrument performs the analysis of the chromatogram and generates a results report on the integral thermal printer.

Performance Data

  Linearity
  The DS5 produces linear results between 4.15 and 14.9% HbA1c.
 
  Within batch precision :
    HbA1c HbA1c
  Number 10 10
  Mean value 5.0 13.2
  SD 0.11 0.26
  CV 2.3% 2.0%
   
  Between batch precision :
    HbA1c HbA1c
  Number 10 10
  Mean value 4.3 7.0
  SD 0.08 0.27
  CV 1.9% 3.9%

Supporting Products

Reagent Kit (300 Tests)

Reagent Kit (75 Tests)
Bi-level HbA1c QC Kit (3 x 0.3 ml level1, 3 x 0.3 ml level2)
 
Technical Data
Data Storage
Storage for 15 chromatograms, results and summary, RS232 Unidirectional Serial Interface Port.
 
 
 
Temperature
Humidity
Operation: 15ºC to 35ºC 10% - 80% non-condensing
Transportation: -5ºC to 40ºC 5% - 80% non-condensing
Voltage: 110/115V or 220/240V:50/60Hz  
Dimension 27cm (w) x 41 cm (d)x 35 cm (h)  
Weight 20 Kg